Clinical Safety; Unexpected Data Loss; Measurements - FujiFilm SonoLite S Serie Guia Del Usuario

Ocultar thumbs Ver también para SonoLite S Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 45

Clinical safety

The following warning was updated.
WARNING

Unexpected data loss

Patient exam calculations on ultrasound workflows can be lost without warning or indication if certain
settings change prior to ending any current (open) studies. See the following caution:
Caution

Measurements

To calculate Stroke Volume (SV) or Stroke Index (SI)
1 Fill in the Height and Weight fields on the patient information form. The BSA is calculated automatically.
(SI only.)
2 Measure from LVOT (2D).
a On a frozen 2D image, press the calcs key.
b From the 2D main menu, select AO/LA.
c From the Calculations menu, select LVOT D.
d Position the calipers.
e Save the calculation.
Clinical safety
FUJIFILM SonoSite does not recommend the use of high-frequency
electromedical devices in proximity to its systems. FUJIFILM SonoSite equipment
has not been validated for use with high-frequency electrosurgical devices or
procedures. Use of high-frequency electrosurgical devices in proximity to its
systems may lead to abnormal system behavior or shutdown of the system.
To avoid the risk of a burn hazard, do not use the transducer with high-frequency
surgical equipment. Such a hazard may occur in the event of a defect in the high-
frequency surgical neutral electrode connection.
Changes made to any of the following settings will cause analysis values in the
current (open) study to be cleared. These settings should not be changed before
ending any open study on the system.
Date and time
OB tables or custom measurements
Cardiac TDI labels
IMT labels
Doppler units
5

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Sonolite edge

Tabla de contenido