9 Examine the system, transducer, and cable for damage, such as cracks or splitting where fluid can enter.
If damage is evident, discontinue use, and contact FUJIFILM SonoSite or your local representative.
10 Prepare the disinfectant for use.
a Choose a high-level disinfectant from the list of approved disinfectants.
High-level disinfectants compatible with SonoSite Edge, and SonoSite S Series transducers
Disinfectant
Compatible transducers
Cidex
C11x, C60x, C60xi, C8x, HFL38x,
HFL50x, ICTx, L25x, L38x, L38xi,
P10x, P11x
Cidex OPA
C11x, C60x, C60xi, C8x, HFL38x,
HFL50x, L25x, L38x, L38xi, L52x
P10x, P21x
1
The P11x transducer is not licensed for use in Canada.
2
The L52x transducer is for veterinary use only.
Refer to the cleaners and disinfection tool available at
more complete list of approved cleaners and disinfectants.
b Check the expiration date on the bottle to ensure the disinfectant has not expired.
c Mix or check that the disinfection chemicals have the concentration recommended by the manufacturer
(for example, a chemical strip test).
d Check that the temperature of the disinfectant is within the manufacturer's recommended limits.
11 Perform a high-level disinfection of the transducer. Use the following procedure:
a Immerse the transducer in a high-level disinfectant solution.
WARNING
Cautions
Do not soak the transducer longer than recommended by the chemical
Do not immerse the transducer connector in any disinfectant solution.
Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a
Cleaning and disinfecting
1
, P21x
High-level disinfectants can cause harm to the patient if not completely removed
from the transducer. Follow the manufacturer's rinse instructions to remove
chemical residue.
manufacturer.
non- recommended disinfecting solution or incorrect solution strength can
damage or discolor the transducer and void the warranty.
Temperature
o
o
25
C (77
F)
o
o
20
C (68
F)
2
,
www.sonosite.com/support/cleaners-disinfectants
Disinfectant Soak Duration
45 minutes
12 minutes
for a