Installed lamp does not match the intended power (watts). Tx450 units require
a 400-watt lamp.
6.2.1.5
Lamp has been used more than 1000 hours. Install new SSX0450 lamp or one used
less than 1000 hours.The main screen always displays lamp hours, and notices are
displayed when a lamp is approaching the end of its useful lamp life. This screen is
displayed when the lamp has exceeded its usable lifetime.
6.2.1.6
In the event of a fan problem the unit will turn off the lamp and display this screen.
Turn off and restart the illuminator. If the fault persists the unit needs servicing
LIT-219 SUNOPTIC SURGICAL
(English)
INCORRECT WATTAGE LAMP INSTALLED
Install correct SSX0450 lamp.
LAMP HAS EXCEEDED USEABLE LIFE
FAN FAULT
®
Rev. -
Date of Revision: 06/21/16