Reset Filter; Error Log - Sunoptic Surgical Titan x450 Manual De Usuario

Ocultar thumbs Ver también para Titan x450:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

6.2.2.4.1
This screen will advise if the air filter is due for replacement. Air filter
replacement every 250 hours of unit operation isadvised. When air filter
replacement is due there will be a flashing signal to the left of
After installing a new filter, touch
the next 250 hours of unit operation.
6.2.2.4.2
The error log functiondisplays error codes useful for repair service. The minus
plus
controls will cycle through the error codes logged.
LIT-219 SUNOPTIC SURGICAL
(English)

RESET FILTER

RESET FILTER

ERROR LOG

®
RESET
FILTER.
to turn off the flashing signal for
Rev. -
Date of Revision: 06/21/16
Page 15 of 104
or

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Tx450d

Tabla de contenido