EN – ENGLISH
Cleaning and Sterilization:
After each use, thoroughly clean and wipe the unit with
antiseptic solution
DO NOT STEAM STERILIZE !
Plastic components may be damaged by heat
Manual Wash
CAUTIONS
• Alkaline and highly acidic detergents and solutions cause
damage to the devices
• Channels and crevices found on this device require
particular attention during cleaning
• Pay special attention to the water quality used throughout
reprocessing Hard water can damage the surface of the
equipment Avoid using hard water Instead use purified
water unless otherwise specified
Limitations on reprocessing
• Repeated processing has minimal effects on these
devices Product life is normally determined by wear and
damage due to use
• It is important to have Integra NeuroSpecialists perform
routine inspections (twice yearly is recommended) See
contact information below
INSTRUCTIONS
Containment/Transportation
• Follow health care facility protocol for safe containment
and transport to the decontamination environment
• It is recommended that devices are cleaned immediately
after use
Figure 3
Cleaning – Manual Equipment: Water, Neutral pH
Detergent, Soft Bristle Brush, Towels
Method
1 Prepare neutral pH enzymatic detergent solution (e g
Endozime® AW Triple Plus with APA (Ruhof), 1:128 ratio)
according to detergent manufacturer's instructions using
lukewarm tap water
2 Prepare equipment for soaking by disassembling
removable parts and loosening connections
3 Rinse equipment in warm water before placing into bath
4 Completely soak equipment in water/detergent solution
for 30 minutes maximum
5 Clean thoroughly with a soft nylon bristle brush NOTE –
If possible, use a disposable brush
6 Rinse in warm purified water until all visible substances
and residual detergent are removed NOTE – Make sure to
give special attention to hard-to-reach areas
7 Thoroughly dry equipment with soft clean towels and
use medically compressed air if needed, to dry channels,
crevices and lumens
8 Inspect the equipment to make sure there is no visible
organic debris or residue from cleaning agent
Repeat process if any soil is detected
Drying
• Products should be dry at this point If wetness or excess
liquid is detected, dry with a soft clean towel
• Medically compressed air can be used if needed
Optional Automatic Wash / Disinfect
CAUTIONS
• Alkaline and highly acidic detergents and solutions cause
damage to the devices
• Channels and crevices found on these devices require
particular attention during cleaning
• Pay special attention to the water quality used throughout
reprocessing Hard water can damage the surface of the
equipment Avoid using hard water Instead use purified
water unless otherwise specified
Limitations on Reprocessing
• Repeated processing has negative effects on these
devices and is not recommended for routine use.
• It is important to have Integra NeuroSpecialists perform
routine inspections (twice yearly is recommended)
See contact information