ENG
9. Charging indication:
When the battery volume <30% the
to be charged in time. When the battery volume <20% the indicator will be twinkled to indicate charge
in time.
10. Overdischarging protection:
When the battery volume <10%, the indicator will power off automatically for avoiding the overdis-
charging.
11. Power off automatically:
In the mode with the automatic power off function chosen, it will display -
consumption if no weighing operation for 3 minutes or more. And it will power off if no weighing ope-
ration for 30 minutes or more.
12. Working time:
The built-in 4Ah battery can supply 15 hours continuous working time with fully charged.
V. Maintenance and Notices:
1. Do not use the indicator under strong sunlight for protecting the using life of it. Keep it on a flat
while using.
2. Do not use it in the places full with dust and strong vibration.
3. Do not overload (including the tare weight) (more than full scale is prohibiten).
4. Do not cleanout the case with strong impregnant (such as benzene, nitric acid oile, etc.).
5. Keep the interior part of indicator away from the water for avoiding the damage to the electrical
components and electronic shock to people.
6. Switch off the power when some problem happened while in operation. Please, send the indicator
to our company for repair, it is not allowed to repair the indicator yourself.
7. The indicator will charge the battery automatically while AC power being connected.
VI. Repair Guarantee & Service (Please send back to our company for repair guaran-
tee and service):
The repair guarantee time is 1 year from the selling date of this indicator, and the probelm should
be happened due to the problem of the indicator itself while in right operation (not due to wrong
operation). We supply the service to the indicator for life. (Battery exclude the repair gurantee).
Note: The repair guarantee not responsible for all problem of the weighing system while our indica-
tor only act as a part of this system, we only take charge of the problem caused from the indicator
itself.
VII. Abnormal display:
weight value > 100% F·S + 9d
OUE r
weight value negative overflow
-OUE r
is on, the battery volume < 30%, please connect with 220V AC power for charging !
is on and
connect with 220V AC power for charging at once !!!
VIII. Parameter setting:
1. Connecting with the load cell well and then doing the parameter set as follows:
2. Press
button for confirmation and then enter next step:
Step Operation
Display
1
Press
18
User's manual. Weighing indicator, counting function TC30
announciator is on, which indicates that the battery needs
(XXXXXX-weight value) twinkled, battery volume < 20%, please
XXXXXX
Notes
In the power off mode, turn on the switch on the back of the
0 -
9
indicator.
0
to decrease power