IX. Capacity calibration:
1. Zero point calibration (Also available on the weighing mode):
Step Operation
Display
1
Press
2
Press
3
Press
2. Capacity calibration
Step Operation
Display
1
Press
2
Press 8888
3
Press
4
Press
Several seconds
5
later
6
Press 3000
7
Press
8
20
User's manual. Weighing indicator, counting function TC30
Notes
- - - - - -
SEt
- - - - - -
nLOAD
CAL 0
00000
Prepare to enter zero calibration.
Must no load on the scale !
0.0
0.00
0.00
Notes
- - - - - -
SEt
Require to input password
- - - - - -
Password = 8888
- - - - - -
SEt
8888
Prepare to enter zero calibration
nLOAD
CAL 0
00000
Must no load on the scale !
- - - - - -
CAL 0
- - - - - -
Prepare to enter full-scale capacity calibration
LOAd
If the loaded weight is not the one as capacity required, the ac-
CAL
tual value can be entered.
600.0
Enter the weight value = 300.0 kg
LOAd
CAL
300.0
The capacity calibration is in process
- - - - - -
CAL
- - - - - -
Several seconds later, it will display the loaded weight value and
300.0
the calibration is finish.
0.00
0.00