User's manual. Weighing indicator, counting function TC30
Step Operation
Display
2
Press
Press 2003
3
4
Press
Press
5
Press
Press
6
Press
Press 3000
7
8
Press
9
Press
30
10
Press
Press 11
11
12
Press
13
Press
14
Press
3. Check the internal code value of A/D transition.
, 9 ,
Press
button, the first line will display the internal code of the present weight value,
press
button to quit.
Notes
- - - - - -
SEt
Require to input password.
- - - - - -
- - - - - -
SEt
Parameter setting code = 2003
2003
1
Enter the parameter-setting mode after password entered.
-d-
Division choise
1
2
1/2/5/0.1/0.2/0.5/0.01/0.02/0.05/0.001/0.002/0.005/10/20/50/
5
100/200/500/0.10/0.20/0.50/0.010/0.020/0.050
0.1
Example: Division d=0.1
Full scale setting
600.0
-FS-
600.0
Example: Capacity (F·S) is 300.0 kg
600.0
-FS-
300.0
Baud filtering parameter choose 00-99, the response speed is
10
faster with small value, but is more stable with big value.
-FLt-
Choose the small one when the weighing condition is better, and
10
choose the big one on the contrary condition.
10
-FLt-
30
Example: baud filtering: FLt=30
One place digit represents the option of automatically power off
10
function or not.
AUtP
One place=0: No automatic power off function.
10
One place=1: Automatic power off.
Tens digit represents zero tracking range (1~9):
1:0.4 d
2:0.8 d
5:2 d
6:2.4 d
also represents the zero range upon power on:
Tens digit=0 not zero when power on
Tens difit≥1 zero rang is 20% F·S when power on
Example: AUtP=11
10
AUtP
(Default AUtP=10)
11
Communication address choice, serial continuous output format
01
Adr=00
-Adr-
Serial output printing Adr=01
01
Press the digit button to input the right address (default Ard=0)
Baud rate choice: 1200
2400
bAUd
2400
Press
to choose
Parameter setting finish
0.0
0.00
0.00
(Default FLt=0)
3:1.2 d
4:1.6 d
7:2.8 d
8:3.2 d
9:3.6 d
2400
4900
9600
(default bAUd=2400)
ENG