EN-4
TSR (Ohms)
350
87 (4-350  cells)
43 (8-350  cells)
35 (10-350  cells)
The T72XW terminal is designed to support both 2mV/V and 3mV/V load cells from the same circuitry. A load cell output
rating selection jumper is not required.
The figure below shows the terminal definitions for the analog load cell terminal strip. Note that when using four-wire load cells,
jumpers must be placed between the +Excitation and +Sense terminals and between the Excitation and Sense terminals
Note: For the standard four-wire cable: If an increase in load results in a decrease in weight display, reverse the signal wires
(+SIG and SIG)

2.5.3 COM1 Serial Port Connections

The COM1 port provides an RS-232 connection for external serial devices.. Make the connections as necessary.
Recommended Maximum Cable Lengths
24 Gauge
(meters/feet)
243/800
60/200
30/100
24/80
Load Cell Termination
Terminal
TxD
RxD
Gnd
COM1 Port Signals
7000 SERIES INDICATORS
20 Gauge (meters/feet)
610/2000
182/600
91/300
60/200
Signal
Transmit RS-232
Receive RS-232
Logic Ground
16 Gauge (meters/feet)
1219/4000
304/1000
152/500
120/400