Figure 4-1 Default Screen
Description:
Sound indicator "
" means the device is on mute status, the user can turn
on the indication sound by short pressing
recall/mute "
" key. During the measurement, over-limit event or probe off
event can activate the alert indication sound.
"
": Memory full icon; if the memory is full, the memory full icon " " appears
on the screen. No display of this icon means the current storing space is not
full. If the memory is full, the data storing will continue in such way the new
record will overwrite the oldest record, so that it's recommended to upload the
stored data into the computer in time.
During the measurement, short time press recall/mute "
" key to turn off (or
resume) the device sound (including pulse beep sound, audible alert and key
click), while the pulse symbol "
" still blinks. For alert sound mute function.
Refer to Section 6.2 for detail of the sound mute function.
Note: The pitch tone of pulse beep (dididi...) is modulated by the SpO2 value,
that means the pitch tone changes when the measured SpO2 changes. The
higher the SpO2 value is, the higher the tone frequency of pulse beep (sound
becomes sharper); The lower the SpO2 value is, the lower the tone frequency
of pulse beep (sound becomes flatter).