I-Tech T-ONE COACH Manual De Uso página 44

Electroterapia
Ocultar thumbs Ver también para T-ONE COACH:
Tabla de contenido

Publicidad

The device is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the device can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter W
0,01
0,12
0,1
0,38
1
1,2
10
3,8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) accordable
to the transmitter manufacturer.
NOTE
1) At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies.
2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
IACER S r
equipment and the device
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d= 1.2 √P
44 de 48
150 kHz to 800
MHz
d= 1.2 √P
0,12
0,38
1,2
3,8
12
800 MHz to 2.5GHz
d= 2.3√P
0,23
0,73
2,3
7,3
23
PG231 Rev 00 17 06 16

Publicidad

Tabla de contenido
loading

Tabla de contenido