Caution: Do not insert the extension or lead connector into the connector block if the
setscrews are not sufficiently retracted. Unretracted setscrews may damage the
extension or lead and prevent the extension or lead from fully seating into the
connector block.
Socket I
(Electrodes 0-7)
Figure 1. Insert the extension or lead connector pins fully into the neurostimulator.
Note: Insert a connector plug (from an accessory kit) into unused neurostimulator socket.
4. For each extension, lead, or plug, fully insert the torque wrench (packaged with the
neurostimulation system) into each self-sealing grommet of the connector block and
tighten each setscrew (Figure 2).
#
Cautions:
Be sure the torque wrench is fully inserted into the self-sealing grommet. If the
torque wrench is not fully inserted, the setscrew may be damaged, resulting in
intermittent or loss of stimulation.
Before tightening setscrews, ensure that the extension or lead connector pins are
inserted into the connector block to prevent damaging the lead or extension.
Verify that each leaf of the self-sealing grommet is closed after the torque wrench
is withdrawn. If fluid leaks through a grommet seal that is not fully closed, the
patient may experience shocking, burning, or irritation at the neurostimulator
implant location, intermittent stimulation, or loss of stimulation.
10 English
37702 2007-12
MA16388A004
Rev A
Medtronic Confidential
NeuroImplant_R01
Socket II
(Electrodes 8-15)
Extension or
lead 1
Extension or
lead 2