Table 1. Operating values for the Model 37702 PrimeADVANCED neurostimulator
Programmable parameter
Number of defined myStim groups
Number of programs per myStim
group
Electrode configuration
Amplitude
Amplitude – upper patient limit
Amplitude – lower patient limit
Pulse width
Pulse width – upper patient limit
Pulse width – lower patient limit
Rate
Rate – upper patient limit
Rate – lower patient limit
SoftStart/Stop
6 English
37702 2007-12
MA16388A004
Rev A
Medtronic Confidential
Operating range and resolution
b
1 to 26
b
1 to 4
2 to 16 electrodes as anode, cathode, or OFF
0 to 10.5 V with 0.05 V or 0.1 V resolution
Tracking limit: programmed value +0 to +4 V (0.5 V
resolution)
Custom limit: programmed value up to 10.5 V (same
resolution as amplitude)
Custom limit: 0 V to the programmed value (same
resolution as amplitude)
60 to 450 µs (10 µs resolution)
Tracking limit: programmed value +0 to +150 µs
(30 µs resolution)
Custom limit: programmed value up to 450 µs (10 µs
resolution)
Custom limit: 60 µs to the programmed value (10 µs
resolution)
2 to 130 Hz (1 Hz resolution to 10 Hz, 5 Hz
resolution from 10 Hz to 130 Hz)
Tracking limit: programmed value +0 to +50 Hz
d
(10 Hz resolution)
Custom limit: 2 to the programmed value (1 Hz
resolution to 10 Hz, 5 Hz resolution from 10 Hz to
d
130 Hz)
OFF, ON: 1, 2, 4, or 8 sec ramp duration
NeuroImplant_R01