Operating Instructions Midwest Stylus
Plus / Midwest
Phoenix ™ /Pro
7 Conditioning
7.4.3

Care of cooling spray nozzle openings

If your tap water is very hard, lime deposits may constrict or completely
block the cooling spray nozzles.
1. Carefully clean the nozzle openings by running a cleaning wire
through them.
2. Let the high-speed handpiece run briefly with cooling spray.
53