Drawing 5
Problem: Error 5~9, the device does not work correctly.
Solution: Remove the battery, await 1 minute and then place it again in the proper
compartment.
Should the error persist, contact your dealer.
Drawing 6
Problem: Temperature exceeds 42.2 2°C - 108°F.
Solution: Verify the integrity of the probe and perform a new measurement.
Drawing 7
Problem: Temperature is lower than 22°C - 71.6°F.
Solution: Verify that the probe is clean and perform a new measurement.
Drawing 8
Problem: The device blocks on self-test.
Solution: Replace the battery
TECHNICAL DATA
Caution: check the instructions for use
BF Type device
IP22: Degree of protection of electrical equipment covers, where the first figure
indicates the degree of protection against penetration by solid foreign bodies
(from 0 to 6) and the second figure the degree of protection against penetration
by liquids (from 0 to 8).
INFRARED FOREHEAD THERMOMETER
Battery:
Display:
Switching-off:
Measuring range:
Resolution:
Operation conditions:
Storage conditions:
Atmospheric Pressure:
Tolerance for body temperature: less than 35°C; +/- 0.3°C
Tolerance for room temperature: from 22°C to 42.2°C; +/-0.3°C
Battery life:
Clinical bias:
Limit of agreements:
Clinical repeatability:
The device complies with the following standards:
ELECTROMAGNETIC COMPATIBILITY
This device complies with current regulations relating to electromagnetic compatibility
(EMC) and has been designed to be used in the home.
Its emissions are extremely small and are unlikely to generate interference with other
20
1 x CR2032LI
liquid crystals
automatic, after one minute, if not used
-22°C and 80°C (-7.6°F and 176.0°F)
0.1°C
10°C to 40°C - RH 15%-85%
-20°C to 50°C, RH <85%
800 - 1013 kPa
from 35°C to 42°C;+/-0.2°C
more than 42°C; +/-0.3°C
<22°C or >42.2°C;
+/- 2°C or +/-4% (highest data)
about 3000 continuous readings
-1.3 ~ -1.5°C.
0.31°C
0.17°C
ASTM E1965-98, EN ISO 80601-2-56, IEC/
EN60601-1-2(EMC), IEC/EN60601-1(safety),
ISO10993, 2011/65/Eu RoHS.