Checking The Extinction Of Light; Use Of Tint Plates - Optika Italy B-380 Serie Manual De Instrucciones

Ocultar thumbs Ver también para B-380 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 44

10.2 Checking the extinction of light

1. Remove the specimen from the light path and
insert 10x objective.
2. Loosen the locking screw of the polarizer ① and
check that the scale is in the "0" position ②. (Fig.
17)
nsert rotatable analyser in the light path, loosen
3. I
the locking screw ③ and put the vibration scale
on 0° ④, then lock the locking screw ③. (Fig.
18)
4. Loosen the polarizer locking screw ① and rotate
the polarizer scale ② to obtain total extinction
(total dark in the eyepieces or "Crossed Nicol
position"). Tighten the screws ①. (Fig. 17)
It may happen that the scale of the polarizer
is not perfectly aligned on the reference mark
but is shifted by one or two notches. This
is not a defect but is due to the mechanical
alignment of the polarizers during assembly.

10.3 Use of Tint plates

Three tint plates are supplied with the microscope:
• l plate (1
order Red)
st
• l/4 plate
"Quartz wedge" plate (Q)
1. Put a specimen on the stage and focus.
2. Insert in the right slot of the Bertrand lens ⑤ one
of the tint plates ⑥. (Fig. 19)
3. In polarized light, inserting one of the plates will
have chromatic effects on the specimen.
Using the l plate (also called 1
specimen will take a magenta tinge.
Using the l/4 plate the specimen will take on a
colour tending to pale yellow.
Using the Q plate the specimen will present a
series of coloured bands that will fade as the
plate is inserted.
order Red) the
st
Page 14
F ig. 17
F
ig. 17
F ig. 18
F
ig. 18
F ig. 19
F
ig. 19

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

B-383pol

Tabla de contenido