Diagnostics Menu - J.P. SELECTA 4002841 Manual De Instrucciones

Extractor de grasas
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES CODIGO 80196 REV B

8.8 Diagnostics menu

The DET GRAS N includes a menu to identify any problem with the measure and temperature controller.
This operation serves to homogenize the temperature of the different equipment internal probes.
• Auto-calibration: This operation allows heater surface temperature calibration against an external ther-
mometer. (It must be provided with a suitable probe for surface measurement).
The process is as follows:
On pressing ENTER, the unit starts for near 30 minutes to stabilize the temperature around 80ºC.
When this point is reached, a repeated beep sounds and the self-calibration process will be over.
• Probes reading: Use this operation to check the temperature probes and the heating elements.
The process is as follows:
Set a working temperature using the Up/Down keys. Validate with ENTER.
Toggle with the ESC key if the readings are shown with (C) or without (N/C) the correction.
Start the unit by pressing ENTER (a "1" is shown on the up right display corner).
Wait until the set temperature stabilizes. The reading of each probe appears on the display.
Stop with ESC. To return to the main menu, keep ESC pressed for a few seconds.
From the main menu, keep ENTER key
pressed for 10 s until the display changes.
PROGRAMMING
DIAGNOSTICS
AUTOCALIBRATION
DIAGNOSTICS
PROBES
J
.P. SELECTA s.a.u.
From the main menu, keep ENTER key pressed for 10s until the diagnos-
tics is shown on the display.
10s
MM:SS
S1 S2 S3 100 avg 0
1 S4 S5 S6
S1 S2 S3 avg 1
1 S4 S5 S6
S1 S2 S3 avg 1
1 S4 S5 S6
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta @jpselecta.es - website: http://www.jpselecta.es
Octubre 2019
(Sujetas a modificaciones sin previo aviso)
1
See process
T 80 ºC
 
Setting the working temperature. Validate the
setting with ENTER.
N/C
Starts the heater to set temperature.
N/C
The unit starts, showing the temperature read for each probe.
S1:
Left corner temperature probe.
C
S6:
Right corner temperature probe.
AVG: Probes average.
N/C: Readings without correction.
C:
Readings with correction.
Pag.: 38