Technical Specifications - ResMed Ultra Mirage Manual De Instrucciones

Ocultar thumbs Ver también para Ultra Mirage:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 45

Technical Specifications

When determining ventilator settings, and in particular the tidal volume for
volume cycled ventilators, make appropriate allowances for mask dead space
and leak.
Unintentional Mask Leak
The Unintentional Mask Leak comprises leak between the cushion-frame and
elbow-frame interfaces.
Pressure
Maximum Leak
(cm H
O)
(L/min)
2
4–20
6
Dead Space
Dead space is the empty volume of the mask up to the elbow.
The dead space of the mask varies according to cushion sizes but is less than
260 mL.
Standards
CE designation in accordance with EC directive 93/42/EEC, class IIa.
EN ISO 14971, ISO 5356-1, ISO 594-1, ISO 10993-1 and ISO 13485.
Operating Pressure
0 to 40 cm H
O
2
Resistance
Drop in pressure measured
at 50 L/min: 0.5 cm H
at 100 L/min: 1.5 cm H
Dimensions
The dimensions of the mask (Medium) are:
H = 200mm, W = 100mm and D = 100mm
O
2
O
2
7

Publicidad

Tabla de contenido
loading

Tabla de contenido