Figure 3a - Proper Insertion of Genius™ Probe Tip into Low Temperature Target
Step 5:
The insertion of the probe tip into the target will automatically be detected and the screen display will change to "Press Genius Scan Button." Press the
Genius™ Thermometer scan button while it is inserted into the Low temperature target. The Genius™ 3 Thermometer will output 3 short beeps as it takes
the temperature of the blackbody target. After the temperature is taken, the next screen will display.
Step 6:
Withdraw the thermometer from the Low temperature target and eject the probe cover. The screen display will change to "Install a new probe cover. Then
insert Genius in High Target Next To Blinking LED." Again, use a new Genius™ probe cover from the cassette and be careful not to point the probe tip at any
object, including hands, fingers or LCD display. Inspect the probe covers to make sure it is fully seated (no space between the cover and the tip base) and
no holes, tears, or wrinkles are present in the plastic film. When the cover is on, insert the probe tip into the High temperature target, which will have a
blinking LED indicator next to it. Make sure that the probe tip is fully inserted into the target well.
Figure 3b - Proper Insertion of Genius™ Probe Tip into High Temperature Target
Step 7:
The insertion of the probe tip into the High target will automatically be detected and the screen display will change to "Press Genius Scan Button." Press the
scan button on the Genius™ Thermometer while it is inserted into the High temperature target. The Genius™ 3 Thermometer will output 3 short beeps as it
takes the temperature of the blackbody target. After the temperature is taken, the next screen will display. Refer to Figure 3b.