INSTRUCTION MANUAL CODE 80234 REV C
CONFIGURATION
CALIBRATION
RETURN
HOME
J
.P. SELECTA s.a.u.
November 2017
ALARM
Alarmas
Alarms
PRINTER
Impresora
Printer
TEMP
Calibración temp.
Temp. calibration
Limit of the Parameters:
SETTINGS:
TEMPERATURE:
CO
:
2
STERILIZATION:
TEMPERATURE:
TIME:
CONFIGURATION:
ALARM:
DATE:
LANG:
PRINTER:
OFFSET TEMP
OFFSET CO
2
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: www.jpselecta.es
(It can be modified without notice)
DATE
Fecha/hora
Date/time
PWD
Acceso servicio tec.
T. assistance access
CO2
Calibración CO
2
CO
calibration
2
Between 10.0 and 50.0 ºC
Between 0,0 and 20,0 %
Between 140 and 180 ºC
Between 1 and 30 minutes
MAXIMUM
TEMPERATURE
Low limit 55 ºC and < SET
MINIMUM
TEMPERATURE
Low limit 10 ºC and > SET
MAXIMUM
% CO
Low limit > SET and 21%
2
MINIMUM
% CO
Low limit 0% and < SET
2
Date and time
Spanish and English
PRINT
TIME
Between 1 h and 24 h
Between -5 and +5 ºC
Between -2 and +2 ºC
LANG
Idioma
Language
EXIT
Salir
Exit
EXIT
Salir
Exit
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Page: 19