7.3 Stating Up
1
Check that the equipment is connected to the power supply, connected to the
CO
cylinder that outlet pressure of the CO
2
0.5 to 1.5 bar by the gas pressure reducer.
2
If working with humidity is needed, fill in the water tank and put it inside, at the
bottom of the chamber.
3
Turn on the main switch (12). During the first 120 seconds, the microprocessor
does not make CO
CO
%. The temperature reading is visible immediately. Nevertheless, during this
2
period, the working parameters can be modified and the incubator can be started
on.
4
To change from STOP mode to RUN mode, press the single button (1) for a while.
The incubator starts to regulate the temperature and the CO
the last used values.
5
When the incubator is in RUN mode, press the single button (1) to change the
incubator to SHOW mode. It shows the set temperature and the CO
tion for about 5 seconds without stopping the operation. Then it goes on with
the normal operation.
6
To stop the incubator, press the single button (1) for at least 5 seconds. The dis-
play will show the stop menu. Select "Yes" by turning the single button (1) and
pressing on it.
7.4 Configuration of Parameters
1
To access the configuration menu, check that the equipment is in STOP mode.
2
Then press the single button (1) for 5 seconds.
3
The menu options are as follows:
TEMP
Ajuste temperatura
Set temperature
TEMP
Temperatura steril.
Steril. temperature
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(Sujetas a modificaciones sin previo aviso) Pag.: 18
readings. During this time, the display shows "--" instead of
2
CO2
Ajuste CO
2
Set CO
2
TIME
Tiempo esterilización
Sterilization time
is adjusted at a pressure between
2
concentration at
2
concentra-
2
EXIT
Salir
Exit
On
Marcha esterilización
Sterilization ON
EXIT
Salir
Exit