4 Maintenance

4.1

General cleaning procedure

4.1.1 General Precautions

Before cleaning always switch off and disconnect from the power supply
Do not autoclave, sterilize, or immerse the instrument or accessory in any fluid
Do not use hard or pointed objects to clean any part of the instrument or accessory
Do not let parts that have been in contact with fluids dry before cleaning
Disinfectant. The use of organic solvents and aromatic oils must be avoided.
Ensure that isopropyl alcohol does not have contact with, any screens on the instruments
Ensure that isopropyl alcohol does not have contact with, any silicone tubes or rubber parts
To prevent a degradation of the material by the Isopropanol it is recommended to irrigate the
water system with distilled water after disinfecting with 70-85% v/v Isopropyl alcohol.
It is recommended that the operator use gloves during the operation of the Air Fx Caloric Irrigator
and the accessories. The gloves are to be changed after each patient, to minimize the contact
points and cross contamination.
Single use components should be replaced after every patient use to avoid potential cross
contamination from patient to patient.

4.1.2 Recommended cleaning agent and frequency

The Air Fx caloric irrigator and the temperature-controlled air is intended for contact with intact skin only.
According to the Spaulding classification used by WHO
product in regard to contamination control. The WHO's recommended level of decontamination for non-
critical devices is cleaning. Disinfection and sterilization are not recommended. However, in case of an
epidemic outbreak, disinfection of the system can be performed.
1.
WHO "Decontamination and Reprocessing of Medical Devices for Health-care Facilities"
Cleaning Agent
The Air Fx Caloric Irrigator is recommended to undergo the regular cleaning procedure with appropriate
cleaning agent. The cleaning agent must be able to remove any foreign material (e. g. , soil,
organic, and microbial contaminants) from the system.
solution like pH neutral detergent as cleaning agent.
Disinfectant
Though the Air Fx Caloric Irrigator is categorized as non-critical device, it is also recommended to
disinfect the system on a regular interval with an appropriate disinfectant to reduce biofilm development.
It is recommended to use 70-85% v/v Isopropyl alcohol as disinfectant agent for the Air Fx Caloric
Irrigator which is also approved by WHO
alcohol 70-85% v/v will also have a mild effect on the materials.
Frequency
The minimum requirement for cleaning and disinfection frequencies are discussed below in detail.
However, the user can decide to improve their cleaning standards with additional cleaning/disinfection
especially during any epidemic outbreaks as per local clinic standards & requirements and WHO
recommendations.

4.2 After each patient

After each examination of a patient, it should be ensured that there is no contamination on the parts in
connection with the patient. Wipe the outside of the otoscope head and replace the single use speculum
after each patient.
Air Fx Caloric Irrigator – Instructions for Use - EN
, it is therefore regarded a non-critical low risk
1
It is recommended to use nonabrasive cleaning
as standard disinfectant agent. Isopropyl
Error! Bookmark not defined. 1
organic, in-
Page 12