Reading Modes - Hach 2100Q Manual Básico Del Usuario

Ocultar thumbs Ver también para 2100Q:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 45
compartment so the diamond or orientation mark aligns with the raised orientation
mark in front of the cell compartment.
1. Push the
2. Insert the 20 NTU
CALIBRATION key to
StablCal Standard and
enter the Calibration
close the lid.
mode. Follow the
Note: The standard to
instructions on the
be inserted is bordered.
display.
Note: Gently invert
each standard before
inserting the standard.
4. Repeat Step 2 and 3
5. Push Done to
with the 100 NTU and
review the calibration
800 NTU StablCal
details.
Standard.
Note: Push Done to
complete a 2 point
calibration.
10 English

Reading modes

1. Push the UP or DOWN key to enter the Reading Options menu.
2. Select Reading Mode to select one of the following options:
3. Push Read. The
display shows
Stabilizing and then
shows the result.
6. Push Store to save
the results.
After a calibration is
complete, the meter
automatically goes into
the Verify Cal mode.
Additional information is
available on the
manufacturer's website.
Option
Description
Normal
The normal mode reads and averages three readings. The
(Default
result is shown after the reading.
setting)
Signal
The Signal Average mode compensates for reading
Average
fluctuations caused by drifting of sample particles through the
light path.
The X-bar icon is shown on the display when signal
averaging is on.
The Signal Average mode measures 12 times and starts to
show the average after three readings. The final result is the
average of all 12 readings.
Rapidly
The Rapidly Settling Turbidity (RST) mode calculates and
Settling
continuously updates the turbidity reading of the sample to a
Turbidity
confidence of 95%, based on the accumulated trend of the
(RST)
real time measured values.
The RST mode is best used on samples that settle rapidly
and continuously change in value. The reading is based on a
correctly prepared sample that is homogeneous at the
beginning of the reading. It is best applied to samples that
are greater than 20 NTU. The sample must be mixed
thoroughly by inversion immediately before inserting it into
the meter.
The target icon is shown on the display when the Rapidly
Settling Turbidity is on.
The Rapidly Settling Turbidity reads and calculates five
readings while showing intermediate results.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

2100qis

Tabla de contenido