Hach 2100Q Manual Básico Del Usuario página 14

Ocultar thumbs Ver también para 2100Q:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 45
7
Troubleshooting
Refer to the following table for common problem messages or
symptoms, possible causes and corrective actions.
Error/Warning
Description
Close lid and push
The lid is open or lid
Read.
detection failed.
Low Battery!
Battery is low.
ADC Failure!
Hardware error causing
reading to fail.
14 English
Detector signal too low! Insufficient light on the
Solution
Overrange!
Make sure that the lid is
closed during reading
and re-read.
• Insert new batteries
Underrange!
• Connect USB/power
module if
rechargeable batteries
are used
Repeat the reading.
8
Error/Warning
Description
180° detector.
Turbidity too high-
caused probably by
calibrating with RapidCal
only.
The measured
absorbance is below the
calibration range.
Solution
• Check for obstructed
light path.
• Check the lamp.
• Calibrate the upper
range.
• Dilute the sample.
Repeat calibration

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

2100qis

Tabla de contenido