AIR FILTER:

The air filter on the side panel of the compressor unit is bright white when
new. With use, it will turn gray, which means it is time to replace it. It must
be replaced immediately if it gets wet or becomes clogged.
To replace the air filter, ensure the compressor is turned off.

FILTER CHANGE

1. DO NOT wash or clean the filter. ONLY use filters supplied by your
distributor and DO NOT operate without a filter.
2. Check the filter every 15 days. Change the filter when it is visibly dirty OR
every 30 days, whichever comes first.
3. Changing procedure (Fig. 12):
a. Remove the filter cover by twisting counter-clockwise and pulling out.
b. Replace the used filter with a new one.
c. Replace the filter cover.