Coltene BioSonic US100R Guía De Uso página 4

Ocultar thumbs Ver también para BioSonic US100R:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 33
4
• Place the dental chair in a supine position for optimal access
to both the maxillary and mandibular arches. This improves
patient comfort and increases clinician visibility.
• Position the patient's head for optimum access to the quadrant
and tooth surface being treated. Evacuate water and debris
with a saliva ejector or high volume evacuator.
B. Ultrasonic scaling procedures
NOTE: Refer to section IV (Maintenance and Infection Control)
of this Owner's Guide for general procedures to follow daily and
between patients.
General operating procedures:
1. Be sure the main dental office water supply valve is open.
2. Turn the unit ON using the ON/OFF button. Verify that the
green LED on the bottom right hand corner of the unit is lit.
3. Set
the
H
O
control
2
open position.
4. Hold the handpiece (without an insert installed) over a sink or
drain. Depress the foot control and flush the water line for at
least 2 minutes.
5. Select the sterilized scaler insert.
6. Lubricate the O-ring with water before placing the insert into
the handpiece. Depress the foot switch pedal while filling the
handpiece with water and slowly turning the insert until it
snaps into place. (Fig. 7)
7. Set the power and H
operating positions. NOTE: Use the lowest effective power
setting possible when scaling. (Scaler inserts should be
sterilized before first use on each patient. See section IV,
part A.)
C. "Turbo" feature
• The "Turbo" button on the bottom left corner of the unit may
be depressed to switch to maximum power, without having to
adjust the preferred power setting. When the "Turbo" mode is
selected, the LED above the button will be lit.
• It is not possible to adjust the power with the power knob as
long as the "Turbo" is on.
• To turn the "Turbo" mode off, depress the button again. The
LED will not be lit.
• The "Turbo" feature is helpful to remove stubborn pieces of
calculus, and should only be used as a temporary mode of
operation.
• It may be necessary to adjust the H
"Turbo" mode to assure adequate fluid delivery to the tip and
the tooth.
D. Helpful hints
• Use the lowest effective power setting possible when scaling.
• If the handpiece becomes hot to the touch, there is either too
little water, too much power or both. Adjust power and/or
water settings accordingly.
• The rubber O-ring on the insert must be lubricated with water
before placing it in the handpiece. Not lubricating the O-ring
will make proper seating more difficult. Fully seat insert by
pushing inward while twisting until seated. DO NOT FORCE.
The handle on the insert should be flush against the handpiece.
Alternatively, lubricating the O-ring with a small amount of
petroleum jelly could prolong the life of the O-ring.
• Check and verify that fluid is reaching the working end of the
insert tip by holding the handpiece over a sink or drain and
activating the unit. Adjust the H
adequate flow is achieved for the selected power setting. A
fine mist should be spraying from the scaler insert tip (Fig. 8).
knob
clockwise
to
O control knobs to your preferred
2
O knob with the unit in
2
O control knob to ensure
2
• If water fails to flow properly, check supply line filter for
clogging. Filters are disposable. Replace if necessary. Do not
attempt to unclog used filters.
• To prevent accidental contact with the lips, tongue, and cheek,
the scaler insert should be activated just before contacting
the tooth, but after the insert has been placed in the
patient's mouth.
• It is best to hold the handpiece in a pen grasp, using a very light
touch when scaling both supra and subgingivally. Too much
pressure of the tip against the tooth surface will decrease the
effectiveness of the insert, as tip vibration is obstructed.
• Periodically check the scaler insert for wear.
• Develop an operating sequence in scaling to ensure
thoroughness, and to minimize adjustments in patient and/or
operator position.
the
fully
E. Patient discomfort can be caused by:
• Excessive pressure. Apply the side of the scaler tip gently to
tooth surface using very little pressure.
• Incorrect tip positioning. Avoid contact of the scaler insert tip
and the tooth. Direct the tip away from root surfaces, and use
side of scaler insert only.
• Not keeping the tip in motion on tooth. The tip should always
be in motion during scaling. Use vertical, horizontal, or
oblique overlapping strokes to increase scaling efficiency and
decrease discomfort.
• If sensitivity persists, alternate treatment of sensitive areas and
less sensitive areas and/or decrease power setting.
IV. Maintenance and Infection Control
A. Ultrasonic Insert Cleaning and Sterilization
Note: DO NOT use chemical disinfectants prior to sterilization or
rapid deterioration of the insert materials may occur.
Remove gross debris by placing the inserts in an ultrasonic cleaner
basket and immersing them in a non-ionic, nonammoniated
cleaning solution. (recommended solutions: Biosonic
UC32). Operate the ultrasonic cleaner in accordance with the
manufacturers directions. Remove the inserts from the cleaner
and dry completely with an air syringe or clean paper towel.
Alternatively , inserts may be carefully hand scrubbed (using a
nylon bristle brush) with detergent and cold water to remove
debris, then rinsed and dried.
Sterilize the Biosonic
sterilizing cycle at 132°C for 15  minutes with a 20 minute
drying time.
Do not use cold sterilization, chemical disinfection, or other
methods of sterilization, including chemical vapor and dry heat
since these methods may severely damage the insert and will
void the warranty.
B. Before each patient
1. Use a sterilized ultrasonic insert for each patient. (Scaler
inserts should be sterilized before first use on each patient.)
2. Clean and disinfect the surfaces of the cabinet, unit, power
cord, and handpiece with Alpet D2. Apply barrier sleeves
to the unit and the handpiece to reduce the risk of cross-
contamination.
3. Flush the water supply lines as discussed previously in
section II.
4. Place the sterilized scaler insert into the handpiece while
maintaining proper asepsis techniques.
BioSonic
Ultrasonic Scaler
®
inserts via a gravity displacement steam
®.
UC30 or
®

Publicidad

Tabla de contenido
loading

Tabla de contenido