6
• During boil-water advisories, this product should not be
operated as an open water system (e.g. connected to a
public water system). A Dental Healthcare Professional should
disconnect the system from the central water source. When
the advisory Is cancelled, flush all incoming waterlines from
the public water system (e.g. faucets, waterlines and dental
equipment) in accordance with the manufacturer's instructions
for a minimum of 5 minutes.
• Use of the Biosonic
®
demineralized tooth structure, hypersensitive areas, veneers,
cast crowns and implants.
Cautions
Observe the following to avoid damaging the de-
vice or impairing performance:
•
Use only inserts manufactured by Coltène.
•
Do not restrict air flow. Provide adequate
ventilation.
CAUTION!
•
Do not subject unit to shock or impact.
•
Do not place the unit on or next to a radiator
or other heat source.
•
Do not attempt to sterilize the handpiece.
Warnings
To reduce the risk of minor burns:
•
Do not operate the unit without water.
•
Use cold (5-25 °C) (41 -77 °F) water only.
WARNING!
•
Do not touch the metal working tip of an
activated insert with bare hands.
To reduce the risk of electric shock:
•
Do not modify the power cord.
•
Do not immerse the unit in water.
WARNING!
•
Use of a Ground Fault Interrupt electrical
outlet is recommended.
To reduce the risk of explosion:
•
Do not use in the presence of flammable
mixtures.
WARNING!
To reduce the risk of water-borne illness:
•
Flush water supply to this device regularly.
WARNING!
US100R Ultrasonic Scaler may irritate
VII. Precautions for Ultrasonic Prophylaxis
Procedures
• Ultrasonic inserts "wear out" with use. Inserts that have 2 mm
of wear lose about 50% of their scaling efficiency.
• If excessive wear is evident, or the insert has been bent, discard
the insert immediately. It may break during use.
• Use retraction to protect the tongue, cheek and lips when the
scaler is in use to avoid contact between the two.
VIII. Order Information
Units:
Catalog No.
Voltage
US100R115M
115V
US100R115T
115V
US100R230CE
230V
US100R230UK
230V
US100R240
240V
Inserts: Frequency 25 kHz
Catalog No.
Description
US1025K
#10 Universal
USG1025K
#10 SuperSoft
US1025KOM
#10 OptiMist™
US1025KSP
#10 Universal Slim
USG1025KSP
#10 Slim SuperSoft
US1025KOS
#10 Slim OptiMist™
USG1025KOS
#10 Slim SuperSoft (Optimist™)
US100025K
#1000 Universal Triple Bend
Inserts: Frequency 30 kHz
Catalog No.
Description
US1030K
#10 Universal
USG1030K
#10 SuperSoft
US1030KOM
#10 OptiMist™
US1030KSP
#10 Universal Slim
USG1030KSP
#10 Slim SuperSoft
US1030KOS
#10 Slim OptiMist™
USG1030KOS
#10 Slim SuperSoft (Optimist™)
US100030K
#1000 Universal Triple Bend
US10OR (OR-alpha)
Replacement O-Ring Gaskets
US100RXF
Water Filter Replacement Kit
IX. Troubleshooting
Generally, check all lines and connections to and from the unit,
a loose plug or connection will often create problems. Check
the settings on the unit's knobs. Although service and repair of
the BioSonic
US100R ultrasonic scaler should be performed by
®
authorized personnel, the following are some basic trouble shooting
procedures that will help avoid unnecessary service calls.
A. Unit will not operate
1. Check the ON/OFF switch making sure it is in the ON position,
and that the detachable power cord is fully seated in the
receptacle on back of unit (ON/OFF indicator light should
be on).
2. Check that the unit's electrical plug is fully seated in an
appropriate AC receptacle, and that the receptacle is active.
3. Disconnect the power cord and check the fuse(s).
BioSonic
Ultrasonic Scaler
®
Frequency
Power
50/60 Hz
100W max
50/60 Hz
100W max
50/60 Hz
100W max
50/60 Hz
100W max
50/60 Hz
100W max
Qty
1
1
1
1
1
1
1
1
Qty
1
1
1
1
1
1
1
1
1
1