Note
1. In the Frequency position, the range conversion is automatic.
Measurement range: 0 ~ 10MHz.
2. Input voltage: 1V rms ~ 20V rms; when the signal frequency becomes higher, the input voltage required by the meter will also increase.
3. For measurement of a small signal with frequency lower than 10 Hz, the signal amplitude must be greater than 2V rms.
- Duty cycle measurement (only for 03057 model)
1. Connect the black probe to "COM" jack and the red probe to "
2. Set the function switch to "Hz%" position.
3. Press and hold "SELECT" button until "%" symbol appears on the screen.
4. Connect the probes to the two ends of the power or circuit to be measured.
5. Take the readings.
Note
When the input signal is not sent, the original reading may be held on the screen. Entering the duty cycle measurement mode again by
pressing "SELECT" button can achieve resetting.
" jack.