Fig. 19: Patency test
WARNING
Pressure admission through the single-use
syringe should be avoided both at the proxi-
mal and the distal end (fig. 20).
Fig. 20: Avoidance of pressurisation
Postoperative valve test
The M.blue has been constructed as a reliably
functioning unit without pump or test function.
The valve test can be performed by flushing,
pressure measurement or pumping.

PRESSURE-FLOW CHARACTERISTICS

Horizontal position
The pressure flow characteristics for the
M.blue differential pressure unit are shown
below for pressure levels 0, 5, 10 and 15 in
the horizontal valve position.
INSTRUCTIONS FOR USE | GB
Fig. 21: Pressure in (cmH 2 O); flow rate in (ml/h)
Pressure-flow characteristics of selected pressure lev-
els of M.blue in the horizontal body position
25