Accuracy Table
Accuracy Table
Accuracy
W������������������������������������������������������
following meanings:
f.s.
(maximum display
value/range)
rdg.
(displayed value)
dgt.
(resolution)
1
AC current 20.00 A/600.0 A (CM4371, CM4372)
Conditions of guaranteed accuracy: After zero adjustment has been performed
Zero-display range:
Coupling type:
Crest factor:
Peak detection time width:
46
The maximum displayable value.
This is usually the name of the currently selected range.
The value currently being measured and displayed on the measuring instrument.
The smallest displayable unit on a digital measuring instrument, i.e., the input value
��������������������������������������
5 counts or less
AC coupling
For the 20.00 A range, 7.5
For the 600.0 A range (300.0 A or less), 3
For the 600.0 A range (greater than 300.0 A and less than or equal
to 600.0 A), 1.5
1 ms or more (Filter off)
HIOKI CM4371A976-00