Accuracy Table
Electrostatic capacity (Measurement value/MAX/MIN/AVG)
(Accuracy guarantee
Range
range)
1.000 µF (0.000 µF to 1.100 µF) 0.001 µF
10.00 µF (0.00 µF to 11.00 µF)
100.0 µF (0.0 µF to 110.0 µF)
1000 µF
(0 µF to 1100 µF)
Auto range movement threshold: 1100 counts or more for upper range, 100 counts or less for lower range.
17
Temperature (same for all models)
Thermocouple:
Instrument reference contact temperature
correction stabilization time:
Overload protection:
Overload current:
68
Resolution Discharge current
10 n/100 n/1 µA±20%
0.01 µF
100 n/1 µ/10 µA±20%
0.1 µF
1 µ/10 µ/100 µA±20%
1 µF
10 µ/100 µ/200 µA±20% ±1.9% rdg.±5 µF
Use the DT4910 Thermocouples (K).
���������������������
error component. [DT4910 error: ±2.5°C (Class 2)]
120 minutes (when ambient temperature is varied abruptly from
60°C to 23°C)
Lower of 1700 V DC/1000 V AC or 2×10
minute)
30 mA or less at steady state, 1.5 A or less at transient state
HIOKI CM4371A976-00
Open
Measurement
terminal
accuracy
voltage
±1.9% rdg.±0.005 µF 2.0 V DC
or less
±1.9% rdg.±0.05 µF
2.0 V DC
or less
±1.9% rdg.±0.5 µF
2.0 V DC
or less
2.0 V DC
or less
Thermocouples (K)
7
V • Hz (energized for 1