Accuracy Table
2
DC current and auto A DC detection 20.00 A/600.0 A (CM4371, CM4372)
Conditions of guaranteed accuracy: After zero adjustment has been performed
Zero-display range:
Coupling type:
Peak detection time width:
DC current (Measurement value/MAX/MIN/AVG)
(Accuracy guarantee
Range
range)
20.00 A (±1.00 A to ±20.00 A)
600.0 A (±1.0 A to ±600.0 A)
Auto range movement threshold: 2000 counts or more for upper range, 180 counts or less for lower range.
DC current (PEAK MAX/PEAK MIN)
Range
(Accuracy guarantee range)
20.00 A (±1.0 A to ±150.0 A)
600.0 A (±10 A to ±900 A)*
* Values of up to ±1500 A are displayed, but accuracy is not defined for display values in excess of 900 A
(which are provided as reference values).
48
5 counts or less
DC coupling
1 ms or more (Filter off)
Measurement accuracy
Resolution
(V����������������������������
0.01 A
±1.3% rdg.±0.08 A
0.1 A
±1.3% rdg.±0.3 A
Resolution
0.1 A
1 A
HIOKI CM4371A976-00
Measurement accuracy
±1.3% rdg.±0.7 A
±1.3% rdg.±7 A
f.)