Maximum heater plate temperature: 185°F (85°C)
Temperature cut-out: 199°F (93°C)
Maximum output humidity: 95% relative humidity (control dial set to maximum setting,
flow rate 50 L/min, ambient temperature of 68°F [20°C], ambient relative humidity of
40%).
Maximum gas temperature: 106°F (41°C)
D
IMENSIONS
Docking station and water chamber: 4.6" x 6.7" x 8.3" (118 mm x 170 mm x 212 mm)
W
EIGHT
Docking station and unfilled water chamber: 1.5 lb. (670 g)
Water capacity to maximum fill line: 415 mL
P
S
OWER
UPPLY
Input range for ResMed S8 flow generator with
110V, 400Hz; 2.5A <140 VA (110 W) (maximum power consumption). Instantaneous
peak power consumption <340 VA.
Actual average power consumption will vary, depending on factors such as
temperature setting, set pressure, the use of accessories, height above sea level and
ambient temperature.
Maximum heater element power: 85 W.
H
C
OUSING
ONSTRUCTION
Docking station: Flame retardant engineering thermoplastic, plated cast aluminium.
Water chamber: Injection molded plastic, stainless steel.
E
NVIRONMENTAL CONDITIONS
Operating temperature: +41°F to +104°F (+5°C to +40°C)
Operating humidity: 10–95% non-condensing
Storage and transport temperature: -4°F to +140°F (-20°C to +60°C)
Storage and transport humidity: 10–95% non-condensing
IEC 60601-1 C
Class II (double insulation)
Type CF
E
LECTROMAGNETIC
Please refer to your flow generator user's manual for details.
Note: The manufacturer reserves the right to change these specifications without notice.
PECIFICATIONS
(H
W
D)
X
X
LASSIFICATION
C
OMPATIBILITY
3i: 100–240V, 50/60Hz;