M.

Error code

If the motor handpiece stops due to an abnormality such as a malfunction, overload,
breakage or misuse, it automatically checks the state of the control unit and detects the
cause of the abnormality, displaying an error code on the LCD panel. If an error code is
displayed, turn on the power again and check whether the same error code is displayed. If
the same error code is displayed, take action by referring to the instructions provided in
the "Check/Remedy" column in the following table.
In case of error, the LCD panel display is as below:
Error code
Error code
At
the time of
charging
During rotation of
the motor
handpiece
At
the time of
calibration
If a fault cannot be eliminated, please contact your distributor.
NVG / F190294X / 06 / 2007 –updated 08/2009
Hold down both buttons of the TORQUE
simultaneously for more than 3 seconds.
"r" for "release" and three digit number
appears on the display
(e.g. 130 for version number 1.30)
Cause
High input voltage in the unit.
Low input voltage in the unit.
High voltage of battery.
High battery current during
charging.
High motor voltage.
High motor load.
High motor load during the
calibration procedure.
Low motor load during the
calibration procedure.
Check/Remedy
Use the original charger.
Otherwise replace the charger.
Use the original charger.
Otherwise replace the charger.
Disconnect the charger.
Otherwise replace the main adapter.
Disconnect the charger.
Contact your distributor.
Switch off the motor.
Contact your distributor.
Check or replace contra angle and motor.
Check or replace contra angle.
Perform calibration with contra angle.