Do not use in the presence of flammable products. The device is not suitable for
use in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
Do not use a bent, damaged, deformed or non-ISO-conforming file. Using such a
file may result in personal injury due to unexpected file breakage or detachment
from the handpiece during rotation.
This motor is equipped with an electronic circuit to limit file breakage. Nevertheless,
files may still break due to higher torque conditions or to fatigue resistance.
Should the leaked battery fluid get into your eyes, immediately wash eyes
thoroughly with clean water and see your doctor. Failure to do so may result in loss
of sight.
Should the battery fluid leak onto skin or clothing, immediately wash the exposed
skin thoroughly with clean water and completely wash away the fluid. Failure to do
so may result in skin complications.
Use of portable phones or other radio frequency (RF) emitting equipment near the
product may cause unexpected or adverse operation.
The product shall not be used adjacent to, or stacked with, other equipment. If
adjacent or stacked use is necessary, the product shall be tested to verify normal
operation in the configuration in which it is being used.
The use of accessories, transducers and cables other than those specified may
result in increased emissions or decreased immunity performance of the product.
The manufacturer claims EMC compliance for the following items: Motor
Handpiece max length 2.0m; Switching Power Supply max length 1.8m
4.

Precautions

 The appliance must only be used with the manufacturer's original accessories.
 If the motor stops or runs too fast, discontinue use and call for assistance.
 Do not introduce any oil into the micromotor.
 The system functions normally in the environment where the temperature is at
10 - 40 ˚ C (50 - 104 ˚ F), humidity at 10 - 85% RH, atmospheric pressure at 500-
1060hPa, and no moisture condensation in the Control Unit. Any use outside these
limits may cause malfunction.
 It is recommended to use rubber dam when performing endodontic treatment.
 Excessive conductive fluid forms a conductive bridge between root canals or with
metallic restorations or crowns. It is recommended to dry the pulp chamber to have
accurate measurement.
 Check the product before use by rotating it, paying attention to looseness,
vibration, noise and temperature (heat generation). If any abnormal condition is
found even slight, at that time, immediately stop use and contact your distributor.
NVG / F190294X / 06 / 2007 –updated 08/2009