Warnings
1. The device must be used exclusively for the intended purpose described.
2. Do not use the device if any component is damaged, loose, detached, or missing. This includes carrier
lenses, telescopes, frame body, temple arms, nose pad, headlight, batteries, cables, and electrical
contacts/plugs. Parts that are broken, missing, distorted or worn should be replaced immediately.
Should such repair or replacement become necessary, ORASCOPTIC OR AN AUTHORIZED SERVICE DEALER
SHOULD PERFORM ALL REPAIRS TO THIS PRODUCT.
3. The use of power adapters, chargers, or batteries other than those provided for use with the device may
result in increased emissions or decreased immunity and may cause the system to be non-compliant
with the requirements of IEC 60601-1-2. Only the power adapter, charger, and batteries that ship with
the device should be used to power the XV1 series.
4. This product maintains compliance with blue light testing standards for safety, set by the International
Electrotechnical Commission (IEC). As with all lighting products, there are risks associated with staring
directly into the light.
• IEC 62471 Risk Group 2 - CAUTION: Possibly hazardous optical radiation is emitted from this product.
Do not stare at the operating lamp. May be harmful to the eyes.
5. Misuse of batteries may cause them to become hot, explode or ignite, and may cause serious injury.
The following safety warnings must be followed:
• Never open or modify the frame, batteries, battery charger, or power adapter.
• Do not use outside.
• Protect from fluids and damp environments.
• Maintain operational environment of 32° to 104°F (0° to 40°C).
• Do not short circuit the electrical contacts.
• Do not bring any part of the device into contact with fire or severe heat over 158°F (70°C).
• If you determine that the device is hot, emits an odor, or changes color or shape, disconnect power
immediately. The device must not be used again.
6. Use of this equipment adjacent to or stacked with other equipment should be avoided because it could
result in improper operation. If such use is necessary, this equipment and the other equipment should
be observed to verify that they are operating normally.
7. Portable radio frequency communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of this device, including
cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
INSTRUCTIONS FOR USE AND CARE
™
Orascoptic
XV1
Technical Support/Customer Care: 1.800.369.3698 | Outside the U.S. or Canada Call Your Local Distributor
™
Series
Laser Warnings
The human eye is extremely sensitive to laser radiation. Failure to observe the following warnings
can result in blindness or other severe and permanent damage to your eyes.
1.
Inspect for damage and fit before each use. Scratched, cracked, or loose fitting laser loupes
or laser inserts may permit the laser beam to reach the eye directly and should not be used.
2.
Ensure the wavelength and power engraved on your laser loupe carrier lenses or laser
insert matches the wavelength of the beam being transmitted by your laser. Different
lasers may require a different type of protective eye wear. If the wavelength at which your laser
operates is not blocked by your laser loupe or laser insert, you must select different eye wear
protection, as your laser loupes or laser insert will not provide protection for your eyes.
3.
Never look directly into a laser beam even when wearing your laser loupe or laser insert.
Your laser loupe is designed to only protect against incidental exposure from stray or diffused
reflection of the laser beam' s energy.
4.
Your laser loupe or laser insert must be worn at all times while operating your laser.
Loupe Prescription Service
If your eye prescription changes, you can send your loupe into our optical laboratory to install your
new prescription. When coordinating the return of your loupe, please allow sufficient time for the
installation of your prescription.
Cleaning and Repair Service
Orascoptic offers a loupe cleaning and repair service that is recommended every two years. Optical lab
technicians will clean and disinfect your telescopes and frame and replace worn-out screws, nose pad,
temple tips, and head strap.
Orascoptic continuously improves its products as a result of customer feedback.
After service, products may reflect improvements in appearance and function.