Electromagnetic Compatibility

Like other electrical medical equipment, SDC Ultra™ requires special precautions
to ensure electromagnetic compatibility with other electrical medical devices. To
ensure electromagnetic compatibility (EMC), SDC Ultra™ must be installed and
operated according to the EMC information provided in this manual.
The SDC Ultra™ has been designed and tested to comply with IEC 60601-1
requirements for EMC with other devices.
Caution
Portable and mobile RF communications equipment may affect
the normal function of the SDC Ultra™.
Caution
Do not use cables or accessories other than those provided with
the SDC Ultra™, as this may result in increased electromagnetic
emissions or decreased immunity to such emissions.
Caution
If the SDC Ultra™ is used adjacent to or stacked with other
equipment, observe and verify normal operation of the SDC
Ultra™ in the configuration in which it will be used prior to using
it in a surgical procedure. Consult the tables below for guidance
in placing the SDC Ultra™
63