Descargar Imprimir esta página

Stryker SDC Ultra Guia Del Usuario página 63

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 353
Error: locating logo.bmp
Error: upgrading software
Application has encountered a critical
error and is shutting down.
No live video on preview
No live video on the monitor
No live video in PIP on preview
No live video PIP on monitor
Ensure that the USB cable is
connected properly. Ensure that the
logo file is present on the USB drive
and it is called "logo.bmp. "
Insert the upgrade dongle in the save
screen.
Please reboot the system.
• Ensure that the camera is on and
check the cable connections.
• Ensure that the proper video source
is selected.
• If the video is routed through
SIDNE®, ensure that SIDNE is
powered on.
• Check the cable connections
• Ensure that the camera is powered
on.
• Ensure that the proper video source
is selected.
• If the video is routed through
SIDNE, ensure that SIDNE is
powered on.
• Check the cable connections
• Ensure that the camera is powered
on.
• Ensure that the proper video source
is selected.
• If the video is routed through
SIDNE, ensure that SIDNE is
powered on.
• Ensure that you are in PIP mode.
• Check the cable connections.
• Ensure that the camera is powered
on.
• Ensure that the proper video source
is selected.
• If the video is routed through
SIDNE®, ensure that SIDNE is
powered on.
• Ensure that the monitor OSD PIP is
enabled via the application.