Appendix A: EMC Information
This device has been designed to meet EMC standards throughout its Service Life without additional
maintenance. There is always an opportunity to relocate your EverFlo device within an environment that
contains other devices with their own unknown EMC behaviors. If you believe your unit is affected by locating
it closer to another device, simply separate the devices to remove the condition.
Appendix A: EMC Information
G
M
'
uidanCe and
anufaCTurer
s
environment specified below. The user of this device should make sure it is used in such an environment.
e
T
Missions
esT
RF radiated emissions
CISPR 11
RF conducted emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker emissions
IEC 61000-3-3
G
M
uidanCe and
anufaCTurer
use in the electromagnetic environment specified below. The user of this device should assure that it is used in such an
environment.
i
T
MMuniTy
esT
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrical Fast Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
d
- e
eClaraTion
leCTroMaGneTiC
C
oMplianCe
Group 1
Class B
Group 1
Class B
Class A
Complies
'
d
- e
s
eClaraTion
leCTroMaGneTiC
ieC 60601 T
esT
l
evel
±8 kV contact discharges
±8 kV contact discharges
±2 kV, ±4 kV, ±8 kV, & ±15 kV air
±2 kV, ±4 kV, ±8 kV, & ±15 kV air
discharges
discharges
±2 kV for power supply lines,
±2 kV for power supply lines,
100 kHz repetition rate.
100 kHz repetition rate.
±1 kV for input-output lines,
NA - Device does not have user
100 kHz repetition rate.
I/O lines that are longer than
3m in length.
±1 kV Differential mode
±1 kV Differential mode
±2 kV Common mode
NA - The device is a Class II
device and does not have earth
connection.
EverFlo / EverFlo Q User Manual
e
: This device is intended for use in the electromagnetic
Missions
e
leCTroMaGneTiC
The device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
The Device is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network.
i
:
MMuniTy
This device is intended for
C
l
oMplianCe
evel
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical home or hospital environment.
Mains power quality should be that of a
typical home or hospital environment.
e
- G
nvironMenT
uidanCe
e
e
leCTroMaGneTiC
nvironMenT
G
uidanCe
-
19