Configuration - WIKA TR36 Manual De Instrucciones

Termorresistencia roscada
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6. Configuration

6. Configuration
Configuration is carried out using a USB interface with a PC via the
model PU-548 programming unit (accessories, order number: 14231581).
EN
The connection with the thermometer is made via the appropriate adapter
cable (accessories: DIN connector, order number: 14005324).
Measuring range, dampening, error signalling, Tag no. and other
parameters can be adjusted (see configuration software).
The measuring range is configurable between -50 ... +250 °C
[-58 ... +482 °F]. The configuration software checks the required
measuring range and will only accept permissible values. Intermediate
values are configurable - the smallest increment is 0.1 °C or 0.1 °F. The
thermometers are delivered configured to customer specifications within
the configurable limits.
Please note:
The measuring range of the thermometer is limited by the application
range of the measuring element, not by the setting range of the
transmitter.
Maximum permissible temperatures:
At the case with transmitter: 85 °C [185 °F]
Class A:
Without neck tube -30 ... +150 °C [-22 ... +302 °F]
With neck tube -30 ... +250 °C [-22 ... +482 °F]
Class B:
Without neck tube -50 ... +150 °C [-58 ... +302 °F]
With neck tube -50 ... +250 °C [-58 ... +482 °F]
18
Easy to use
LED status display
Compact design
No further voltage supply is needed for either
the programming unit or for the transmitter
(replaces programming unit model PU-448)
WIKA operating instructions model TR36

Publicidad

Tabla de contenido
loading

Tabla de contenido