Enter
Used to confirm a numerical entry (value, bin number or program store
number).
Store/Recall Keys
The following keys are used to store and recall set—ups:
Store
Stores the complete set—up, including the set binning values, in non volatile
memory.
Recall
Recalls up to nine previously stored set—ups.
Simple Pass / Fail Sorting
To set up simple pass/fail sorting, first select the measurement type to be
made, i.e. R+Q, L+Q, C+D or C+R. Set the measurement frequency and select
series or parallel measurement as required.
Note: Binning cannot be set with the bridge in Auto mode.
Bin Selection
Press the Bin No. key to enter set—up mode. Successive presses of the
Bin No. key will step the display through the options of binX (where X is the bin
number), CLEAr? (clear the selected bin), CLEAR ALL? (clear all bins) and
End? (exit bin set—up mode). The first press of Bin No. will enter the option
sequence where it was last exited; it may be necessary to press the key several
times to get to the desired option.
If any previous binning information needs to be cleared select CLEAr?
with the Bin No. key and press Enter; the display should show the message
CLEAr donE and then binX in the right—hand display, ready for the next step.
If all bins are to be cleared select CLEAr ALL? and follow a similar procedure.
For simple pass/fail sorting, bin 0 must be used. The other bins (1 to 7
inclusive) should be 'closed' by setting their limits to zero; alternatively, and
easier, all the bins can be cleared by using CLEAr ALL? before bin 0 is set.
Bin 8 can be used to set limits for the minor parameter (Q, D, or R); parts that
fail these limits fall into bin 8. Parts that fall into neither bin 0 nor bin 8 fall into
bin 9, the general fail bin.
Press Bin No. until binX shows in the display. Press 0 to select bin 0;
bin0 should show in the right—hand display.
Page 24
MZ-805 L/C/R meter