The following maintenance and/or sterilization
instructions should be carried out before using
the accessories supplied with the device for the
first time.
Avoid using cleaning agents and disinfectants
containing inflammable substances. If this is
unavoidable, ensure that the product has
evaporated and there is no combustible product
on the device and accessories before operation.
6. 1 SUPRASSON NEWTRON HANDPIECE OR
SUPRASSON NEWTRON LED HANDPIECE
MAINTENANCE
After each use, the water circuits of the whole
unit comprising the S
should be rinsed with clean water for 20 to 30
seconds.
Disconnect the handpiece from its cord before
cleaning, disinfection and sterilization.
For sterilization of the handpiece, please refer to
the specific instructions for the product.

6. 2 TIP MAINTENANCE

WARNING: during sterilization, metal parts of
different types must not touch each other. Any
contact will lead to the formation of
electrolytic
couples
deterioration. In order to avoid this happening,
place the objects to be sterilized in a cloth or
sterilizing bag.
a) Lifetime
Tip shape and weight are the determining
characteristics for obtaining maximal efficiency
of the ultrasonic generator.
Careful monitoring of these two characteristics
by the user will ensure that optimum
performance of the device is maintained.
Consequently, you are strongly advised not to
N
®
and the tip
UPRASSON
EWTRON
causing
localized
alter the structure of the tips by filing or twisting
them or modifying them in any other way.
Similarly, tip aging leads to alteration of its
characteristics through normal wear.
Always replace a tip that has been damaged by
wear or by accidental impact (dropped,
deformed, etc.).
Use the T
C
to check scaler tip wear. Replace
IP
ARD
routinely-used tips at least one to three times a
year.
b) Pre-disinfection / Chemical Cleaning
To be carried out immediately after treatment,
wearing thick gloves.
Immerse in an ultrasonic bath (enzymatic,
quaternary ammonium base detergent /
disinfectant solution), carefully respecting the
concentration and contact time recommended by
the manufacturer of the solution.
Use a solution that bears the CE marking or that
is in compliance with any standard required by
the national regulations.
Rinse under running water for at least 30
seconds.
c) Mechanical / Chemical Cleaning
To be carried out immediately after pre-
disinfection / chemical cleaning, wearing thick
gloves.
Scrub the products in a new bath of enzymatic,
quaternary ammonium base detergent /
disinfectant solution, for at least 30 seconds, and
in any case until all remaining traces of
contamination have been removed.
Rinse again under running water for at least 30
seconds.