Electromagnetic Compatibility; Electromagnetic Emissions - Satelec Acteon P5 Newtron Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 36
VIII - ELECTROMAGNETIC COMPATIBILITY
Important:
The power cable, the handpiece cable and the footswitch cable must be kept apart.
The device requires special precautions to be taken with regard to electromagnetic compatibility.
It must be installed and prepared for use as described in chapter 4.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere
with the device.
The recommended separation distances in this paragraph must therefore be complied with.
The device must not be used near or on top of another device.
If this cannot be avoided, its operation under the conditions of use must be checked beforehand.
The use of accessories other than those specified or sold by SATELEC as replacement parts may have
the consequence of increasing the emissions or decreasing the immunity of the device.

8. 1 ELECTROMAGNETIC EMISSIONS

The device is intended for use in the electromagnetic environment specified in the table below.
The user and/or installer must ensure that the device is used in such an environment.
Emission test
RF emission - CISPR 11.
RF emission - CISPR 11.
Harmonic current emission
IEC 61000-3-2.
Voltage fluctuation and flicker
IEC 61000-3-3.
Compliance
The device uses RF energy for internal operation.
Group 1
Therefore, its radiofrequency emissions are very low and are not likely
to cause any interference in nearby equipment.
Class B
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low voltage
Class A
power supply network that supplies buildings used for domestic
purposes.
Complies
Electromagnetic environment - guidance
15

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

P5 newtron led