- Before Turning On the Equipment
1º - Check if the equipment is properly installed in accordance with the
instructions in chapter 6 (Installation) of the User Manual Dental Line;
2º - Turn On the air compressor that supplies the equipment;
3º - Open the main water supply valve of the equipment;
4º - Turn On the equipment's power circuit breaker.
- Turning On the Equipment
1º - Turn On the On/Off switch (4);
2º - Activate the "Seat Up" command on the pedal (section 5.3) until the seat
reaches its maximum height limit;
3º - Open the protection cover (3);
4º - Close the pressure valve (1);
5º - Make sure the water tanks (2) are filled with water. If necessary, fill it with
drinkable water up to the limit indicated in each tank and reattach them
again. If you wish to place prophylactic products (low concentration);
6º - Open the pressure valve (1) and check if the water tanks (2) are well
attached and free of leaks;
7º - Close the protection cover (3);
8º - Before starting to use the equipment, check the operation of all available instruments and controls and
perform cleaning and sterilization of the instruments (chapter 7 of the User Manual Dental Line - Cleaning
and Disinfection).
Do not move the chair with the protection cover open.
5.2.1 - Delivery System
1 - Pressure Valve
2 - Water Tanks
3 - Protection Cover
4 - Parts Identification
5 - Equipment Operating and Description
5.2.2 - Electric Panel
4 - On/Off Switch
5 - Fuse Holder
6 - Digital Voltmeter
7 - Emergency Stop Button (Optional)
7
A - Delivery System
B - Electrical Panel
C - Remote Foot Pedal
D - Working Table
E - Assistant Module
F - Operating Light
G - Dental Chair
H - Water Unit