Implementation of the provision Iron, total (Fe, Mo) in the
presence of molybdate with Vario Powder Packs
Select the method on the device
50 ml
Put 50 ml sample in 50 ml
measuring cylinder
Blank
Prepare two clean 24 mm
vials. Mark one as a blank.
25 ml
Put 25 ml prepared
sample in 25 ml measuring
cylinder.
EN   Method Reference Book   1.0
Add Vario (Fe in Mo) Rgt
1 powder pack.
10 ml
Blank
Fill blank with 10 ml
prepared sample.
Add Vario (Fe in Mo) Rgt
2 powder pack.
Iron in Mo PP (224) / M224
Stopper the mixing
cylinder. Swirl around to
dissolve the powder.
Close vial(s).
Stopper the mixing
cylinder. Swirl around to
dissolve the powder.