Iron in Mo PP (224) / M224
EN
Press the ENTER button.
Blank
Place blank in the sample
chamber. • Pay attention to
the positioning.
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
The result in mg/l Fe appears on the display.
48
3 min
12
11
1
10
2
9
3
8
4
7
5
6
Wait for 3 minute(s)
reaction time.
Zero
Press the ZERO button.
Test
Press the TEST (XD:
START)button.
EN Method Reference Book 1.0
10 ml
24 mm
Put 10 ml sample in the
sample vial.
Remove the vial from the
sample chamber.