Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/l Ozone; mg/l total chlorine appears on the display.
Implementation of the provision Ozone, in absence of chlorine
with tablet
Select the method on the device
In addition, choose the test: without Chlorine
10 ml
Fill 24 mm vial with 10 ml
sample.
Zero
Press the ZERO button.
EN   Method Reference Book   1.0
Close vial(s).
Remove the vial from the
sample chamber.
Ozone T / M300
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Empty vial except for a few
drops.
EN
71