Descargar Imprimir esta página

Stryker Secure Connect Manual De Operaciones/Mantenimiento página 21

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

8. Using diagonal pliers, cut the cable tie (3000-300-114) (AK) that secures the 37-pin cable (M) to the cable routing tray
(D) (Figure 9). Discard the cable tie.
9. Move the 37-pin cable (M) and cable routing tray (D) up to remove the battery (700000341245) (V) from the body (C).
10. Unplug the two battery cables and discard the battery.
N N o o t t e e - - When you reinstall, make sure that the red wire connects to positive (+) and the black wire connects to negative
(-) on the battery.
11. Reverse steps to reinstall.
12. Verify proper operation before you return the product to service.
5212-009-205 Rev AA.1
F F i i g g u u r r e e 8 8 – – R R e e p p l l a a c c i i n n g g t t h h e e b b a a t t t t e e r r y y ( ( l l e e f f t t ) )
F F i i g g u u r r e e 9 9 – – R R e e p p l l a a c c i i n n g g t t h h e e b b a a t t t t e e r r y y ( ( l l e e f f t t ) )
17
EN

Publicidad

loading

Este manual también es adecuado para:

521200380100