Descargar Imprimir esta página

Stryker Secure Connect Manual De Operaciones/Mantenimiento página 22

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

R R e e p p l l a a c c i i n n g g t t h h e e b b a a t t t t e e r r y y ( ( 3 3 7 7 - - p p i i n n c c o o n n n n e e c c t t o o r r , , r r i i g g h h t t s s i i d d e e ) )
T T o o o o l l s s r r e e q q u u i i r r e e d d : :
#2 Phillips screwdriver
#1 Phillips screwdriver
Needle nose pliers
Straight pick
P P r r o o c c e e d d u u r r e e : :
1. Unplug the power supply from the S S e e c c u u r r e e C C o o n n n n e e c c t t .
2. Using a #1 Phillips screwdriver and needle nose pliers, disconnect the nurse call communication cable from the S S e e c c u u r r e e
C C o o n n n n e e c c t t .
3. Using a #2 Phillips screwdriver, remove the two screws (G) that secure the S S e e c c u u r r e e C C o o n n n n e e c c t t (F) to the mounting plate
(D) (Figure 3). Save the two screws.
4. Using a straight pick, remove and discard the label (521200380902) (C) that covers the O O N N / / O O F F F F switch (B) (Figure 2).
5. Turn the O O N N / / O O F F F F switch (B) to the O O F F F F position (Figure 2).
6. Using a #2 Phillips screwdriver, remove the ten screws (AC) that secure the front cover to the body (Figure 7). Save the
screws and front cover.
7. Unplug the two battery cables and discard the battery (700000341245) (V).
N N o o t t e e - - When you reinstall, make sure that the red wire connects to positive (+) and the black wire connects to negative
(-) on the battery.
8. Reverse steps to reinstall.
9. Verify proper operation before you return the product to service.
EN
18
5212-009-205 Rev AA.1

Publicidad

loading

Este manual también es adecuado para:

521200380100