I I n n s s t t a a l l l l a a t t i i o o n n
I I n n s s t t a a l l l l i i n n g g S S e e c c u u r r e e C C o o n n n n e e c c t t
T T o o o o l l s s r r e e q q u u i i r r e e d d : :
•
#2 Phillips screwdriver
•
#1 Phillips screwdriver
•
Straight pick
•
Tape measure
•
Level
•
Pencil
•
Tools required for hospital supplied fasteners
P P r r o o c c e e d d u u r r e e : :
1. Record the S S e e c c u u r r e e C C o o n n n n e e c c t t BBID ( Serial number location (page 5)) and room number/patient position on the S S e e c c u u r r e e
C C o o n n n n e e c c t t association form (page 37).
2. Using a straight pick, configure the SB1 and SB2 dip-switches (A) to match the nurse call system and nurse call
communication cable (Figure 2).
C C A A U U T T I I O O N N - - Always match the dip-switches on SB1 and SB2 to the connected product configuration to avoid the risk of
head wall damage.
N N o o t t e e - - To confirm dip-switch configuration, contact Stryker customer service or technical support ( Contact information
(page 5)).
3. Using a straight pick, turn the S S e e c c u u r r e e C C o o n n n n e e c c t t O O N N / / O O F F F F switch (B) to the O O N N position (Figure 2).
4. Install the two supplied IPX labels (C) over the dip-switch and the O O N N / / O O F F F F cutouts located on the back of the S S e e c c u u r r e e
C C o o n n n n e e c c t t (Figure 2).
5. Using a tape measure and pencil, mark the center of the product location between 4 in. - 5 in. (10.2 cm - 12.7 cm) from
the floor (Figure 3).
N N o o t t e e - - The marked mounting location on the wall must be within 0.5 in. (1.3 cm) left or right from the center line of where
the product will be positioned for the S S e e c c u u r r e e C C o o n n n n e e c c t t to work properly.
EN
6
B
C
F F i i g g u u r r e e 2 2 – – S S w w i i t t c c h h c c o o n n f f i i g g u u r r a a t t i i o o n n
A
5212-009-205 Rev AA.1